A solid-state microdosimeter for dose and radiation qualty monitoring for astronauts in space
Peracchi, Stefania; James, Benjamin; Tran, Linh T.; Bolst, David; Prokopovich, Dale A.; Davis, Jeremy; Guatelli, Susanna; Petasecca, Marco; Lerch, Michael L.F.; Matsufuji, Naruhiro; Kok, Angela; Povoli, Marco; Rosenfeld, Anatoly B.
Journal article, Peer reviewed
Accepted version
View/ Open
Date
2019Metadata
Show full item recordCollections
- Publikasjoner fra CRIStin - SINTEF AS [5647]
- SINTEF Digital [2395]
Original version
10.1109/TNS.2019.2943597Abstract
This paper presents a study of the response of the silicon on insulator (SOI) microdosimeter with 3-dimensional (3D) sensitive volumes (SV) to 400 MeV/u 16O and 500 MeV/u 56Fe ions mimicking Galactic Cosmic Rays outside and inside the International Space Station (ISS). An average quality factor (̅) and the dose equivalent (H) of the radiation field were obtained experimentally and the results were compared with GEANT4 simulations.
Description
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TNS.2019.2943597, IEEE Transactions on Nuclear Science