Microdosimetry with a 3D silicon on insulator (SOI) detector in a low energy proton beamline
Samnøy, Andreas Tefre; Ytre-Hauge, Kristian; Malinen, Eirik; Tran, Linh; Rosenfeld, Anatoly; Povoli, Marco; Kok, Angela Chun Ying; Summanwar, Anand; Røhrich, Dieter
Peer reviewed, Journal article
Published version
Date
2020Metadata
Show full item recordCollections
- Publikasjoner fra CRIStin - SINTEF AS [5967]
- SINTEF Digital [2556]
Original version
10.1016/j.radphyschem.2020.109078Abstract
An accurate description of the radiation quality of proton beams is a precondition to increase our understanding of radiobiological mechanisms and to develop accurate biological response models for radiotherapy. However, there are few detectors capable of measuring microdosimetric quantities with high spatial resolution along the entire Bragg curve due to the rapid increase in stopping power at the Bragg peak (BP) and distal dose fall-off (DDF). The aim of this work was to measure the microdosimetric spectra along the Bragg curve in a low energy proton beamline used for radiobiological experiments with a novel 3D silicon-on-insulator (SOI) “mushroom” microdosimeter.