Validation of linear energy transfer computed in a Monte Carlo dose engine of a commercial treatment planning system
Wagenaar, David; Tran, Linh T.; Meijers, Arturs; Marmitt, Gabriel Guterres; Souris, Kevin; Bolst, David; Povoli, Marco; Kok, Angela; Traneus, Erik; Van Goethem, Marc-Jan; Langendijk, Johannes A.; Rosenfeld, Anatoly B.; Both, Stefan
Journal article, Peer reviewed
Accepted version
Permanent lenke
http://hdl.handle.net/11250/2636404Utgivelsesdato
2019Metadata
Vis full innførselSamlinger
- Publikasjoner fra CRIStin - SINTEF AS [5470]
- SINTEF Digital [2334]
Originalversjon
10.1088/1361-6560/ab5e97Sammendrag
The relative biological effectiveness (RBE) of protons is highly variable and difficult to quantify. However, RBE is related to the local ionization density, which can be related to the physical measurable dose weighted linear energy transfer (LETD). The aim of this study was to validate the LETD calculations for proton therapy beams implemented in a commercially available treatment planning system (TPS) using microdosimetry measurements and independent LETD calculations (Open-MCsquare (MCS)). The TPS (RayStation v6R) was used to generate treatment plans on the CIRS-731-HN anthropomorphic phantom for three anatomical sites (brain, nasopharynx, neck) for a spherical target (Ø=5 cm) with uniform target dose to calculate the LETD distribution. Measurements were performed at the University Medical Center Groningen proton therapy center (Proteus Plus, IBA) using a + -probe utilizing silicon on insulator microdosimeters capable of detecting lineal energies as low as 0.15 keV/m in tissue. Dose averaged mean lineal energy ̅̅̅ depth-profiles were measured for 70 and 130 MeV spots in water and for the three treatment plans in water and an anthropomorphic phantom. The ̅̅̅ measurements were compared to the LETD calculated in the TPS and MCS independent dose calculation engine. D⋅̅̅̅ was compared to D⋅LETD in terms of a gamma-index with a distance-toagreement criteria of 2 mm and increasing dose difference criteria to determine the criteria for which a 90% pass rate was accomplished. Measurements of D⋅̅̅̅ were in good agreement with the D⋅LETD calculated in the TPS and MCS. The 90% passing rate threshold was reached at different D⋅LETD difference criteria for single spots (TPS: 1% MCS: 1%), treatment plans in water (TPS: 3% MCS: 6%) and treatment plans in an anthropomorphic phantom (TPS: 6% MCS: 1%). We conclude that D⋅LETD calculations accuracy in the RayStation TPS and open MCSquare are within 6%, and sufficient for clinical D⋅LETD evaluation and optimization. These findings remove an important obstacle in the road towards clinical implementation of D⋅LETD evaluation and optimization of proton therapy treatment plans.
Beskrivelse
Accepted Manuscript is “the version of the article accepted for publication including all changes made as a result of the peer review process, and which may also include the addition to the article by IOP Publishing of a header, an article ID, a cover sheet and/or an ‘Accepted Manuscript’ watermark, but excluding any other editing, typesetting or other changes made by IOP Publishing and/or its licensors” This Accepted Manuscript is © 2019 Institute of Physics and Engineering in Medicine.